Home Company Info Product List Contact Details
Homeproduct Directory

JK300 Monochromator:(Manual Counter)

JK300 Monochromator:(Manual Counter)
company Beijing Jike Instrument Co.,Ltd
Categories Acrylic Adhesives
Update2011-03-31
Original RegionChina
JK300 Monochromator:(Manual Counter)

JK300 Monochromator:(Manual Counter)


Model:JK300
As monochromatic light source,JK300 Monochromator can make qualitative and quantitative analysis to transparent substance and solution as well as measuring to the apectral components of polychromatic light source,Its application will be further extended if necessary accessories and testing units are equipped .Inrecent years it is widely used in physics in physics,chmistry,metallurgy,agriculture,environmental protection,building material and medicine departments as indispensable testing instrument in production,scientific research and teaching
Technical Parameters
Wavelength Range300nm~1000nm(extensible to 190nm~20m)
Wavelength Accuracy0.2nm(1200line/mm grating)
Wavelength Repeatability0.1nm(1200line/mmgrating)
Focal Length300mm
Relative ApertureD/f=1/6
Resolution0.2(1200line/mm grating,measured by means of spectral line half width in compliance with intermatonal regulations)
Linear Dispersion Reciprocal2.7nm/mm(1200line/mm grating)5.4 nm/mm(600line/mm grating
Stray Light1×10-3(He-Ne lasermeasured at 632.8±8nm)
SlitWidth:0.01~3mm continuously adjustableHeight: 5 type:2mm4mm6mm8mm10mm GratingStandard Configuration:plane grating:1200line/mm score:50mm×50mmblaze wavelength:b=500nm(other custom grating available).
Dimensions385× 260 ×190)mm
JK300-based system can help the user with three different filter. Filter the use of 350nm-600nm; 600nm-1000nm; 1000nm-1500nm. When you choose to use more than 1500nm wavelength of the grating, to be equipped with the appropriate cut-off-infrared filters.
The specific choice is as follows:
Ordinary optical glass
Transmission range 300nm-2500nm
Quartz material
Transmission range 200nm-3500nm
As of self-filter
Wavelength> 3500 nm
1#filter
The use of band
350nm-600nm
<300nmT<2%>400nmT>80%
2#filter
The use of band
600nm-1m
<540nmT<2%>590nmT>80%
3#filter
The use of band
1m-1.5m
<750nmT<2%>1000nmT>80%
Instruments used in the most reflective diffraction grating is blazed grating. Blazed diffraction grating has control of its energy in the spatial distribution of properties, it is particularly valuable in use. I offer the following types of companies grating:
Grating described a few lines (L/mm)
Wavelength shine p(nm)
Use Range(nm)
1200
300
200-600
1200
500
330-1000
600
500
330-1000
600
750
500-1500
300
1250
800-2500
300
2000
1300-4000
150
2000
1300-4000
150
4000
2600-8000
100
1500
1000-3000
100
2000
1300-4000
JK300-grating monochromator -2 can be equipped as follows:
Receiver Series
Light Series